Welcome to BLX foreign trade company!

Email

xiaoniuyisheng123@163.com

WhatsApp

+8613676902996

Inspection and repair of damaged B-ultrasound lens and lens failure in animals

2024-09-26 13:58:46 Visits:

Repair of damaged animal B-ultrasound crystal, fault phenomenon: black shadows are mixed in the animal B-ultrasound image detected at one time, and there are black strip interference. Analysis and repair: Most animal B-ultrasound probes use piezoelectric crystals. After long-term use, the crystals will age naturally. Irregular use or external force such as falling or bumping will cause damage to the probe crystals. After the crystal is damaged, the image detected by the animal B-ultrasound detector will show signal attenuation, black shadows, black bars, interference, defects and other blind areas in the image, which will make it impossible for doctors to diagnose, affect the quality of diagnosis, and even cause misdiagnosis.

This type of fault is caused by a short circuit in the animal B-ultrasound probe chip. The short circuit of the chip can be judged by measuring the power supply voltage of the chip. If the power supply voltage of a chip or chip group is lower than that of others, it can be judged that there is a short circuit in the chip or chip group. Replace the corresponding chip or chip group to restore the original characteristics of the veterinary B-ultrasound probe and eliminate the interference.

Repair of animal B-ultrasound lens failure, fault phenomenon: vertical dark stripes appear in the image of the veterinary B-ultrasound display. The vertical dark stripes on the probe were squeezed by hand, and the vertical dark stripes were significantly weakened. The vertical dark stripes disappeared when the same model animal B-ultrasound probe was replaced. It was preliminarily judged that the acoustic lens layer and the matching layer might be debonded.

Analysis and maintenance: The acoustic lens is a colloid substance on the upper part of the veterinary B-ultrasound probe that contacts the animal body. After long-term use, the acoustic lens will be naturally worn, scratched, cracked, corroded, debonded, bubbled, etc., especially the corrosion caused by the use of inferior coupling agent entering the lower layer, which will not only cause artifacts or blind spots, but also lead to misdiagnosis by doctors. In severe cases, it will damage the animal B-ultrasound probe crystal or cause leakage, endangering the patient's physical safety and causing harm to animals and machines.



tags: animal B-ultrasoundveterinary B-ultrasound probe
Text link:https://www.bxlultrasound.com/faq/296.html
animal B-ultrasound veterinary B-ultrasound probe

Leave Your Message


Leave a message